Display
LCD display: 320 x 240 monochrome crystal display with backlight (touch panel type)
Display details: UV absorbance, VIS absorbance, UV-VIS absorbance (Absorbance can be changed to absorption constant (SAC: m-1), Converted COD level, Converted turbidity level
Communication
Interface: RS-232C Compatible Communication speed: 19200 bps
Analogue Outputs
3 outputs (standard)
Up to 3 can be chosen from UV absorbance, VIS absorbance, UV-VIS absorbance, Converted COD level, and Converted TURB level outputs
DC 4 – 20 mA, or DC 0 – 16 mA, isolated output (no isolation between each channel), maximum resistance load: 600Ω
Contact Inputs
2 inputs (standard)
Sample failure float switch input, time correction input
No-voltage contact input (can connect to open collector), isolated input
On resistance: Maximum 100Ω, Open voltage: DC 26 V, Short circuit current: 13 mA
Alarm and Status Outputs
6 outputs (standard)
Up to 4 can be chosen from power failure, maintenance (standard and fixed), batch alarm, COD maximum limit alarm, TURB limit alarm (standard), light source error, sample failure, cleaning motor error, and analyzer error outputs
Power failure…occurs when power fails Total alarm…cleaning motor error, light source error, analyzer error Maintenance…occurs when entering maintenance or correction mode, or when the maintenance switch has been turned on
Dry contact output, NO contact
Contact rating: AC 125 V 0.3 A, DC 30V 1 A (with resistance load) Each output has an independent COM interface
Data Logger
Values of measurement items are stored in the main unit”s data memory. Data can also be transferred to a Compact Flash card.
Intervals 1 minute or 1 hour
Capacity 1 minute intervals 0 for 10 day”s data, 1 hour intervals 0 for 1 year” s data